Assessment of Interobserver and Intraobserver Variabilities in the Measurements of Anteroposterior (AP) and Lateral (LAT) Diameters on Paediatric CT images for Size-Specific Dose Estimation.
Loading...
Files
Date
Journal Title
Journal ISSN
Volume Title
Publisher
Asian Journal of Medicine and Biomedicine
DOI
Abstract
Both interobserver and intraobserver variability evaluations are important to validate the data reliability
and ensure consistency in measurements. to assess the variability in anteroposterior (AP) and lateral (LAT)
diameter measurements of the head and abdomen for determination of effective diameter among paediatric
population. The measurements were conducted by various observers. Fifty CT images of paediatric head
and abdomen were randomly selected from from five different age groups of paediatric patients, all aged
below 18 years old. All images were acquired using the standard exposure protocol for head and abdomen
CT scanning. The anteroposterior (AP) and lateral (LAT) diameters were measured on selected CT images
of the head and abdomen at specific levels. To ensure measurement consistency, diameters were
consistently measured at defined positions: the middle slice for the head and a position between the
kidneys or the second lumbar vertebra for the abdomen. All diameter measurements were conducted by
three observers and repeated three times. Statistical analysis, including the intraclass correlation
coefficient (ICC) and Bland-Altman method for intraobserver and interobserver agreement, were employed
to assess measurement consistency. The findings showed excellent reliability in all measurements (LCC
values > 0.9) for both interobserver and intraobserver. The variability between student, physicist and
radiologist was high. The deviation of observer measurements between all observers were also significant
for AP and LAT diameters of head and abdomen of the selected CT images. The findings resulted in high
reliability, showing strong agreement among measurements and low variability in measurements among
all observers and amongst all participating centres. The variability between the principal investigators was
lower, but still substantial. The deviation of observer measurements from the experience observer
(reference) were also significant for all CT parameters. We found excellent reliability for both intra- and
interobserver measurements.
Description
Keywords
Citation
Interobserver Variability; Intraobserver Variability; Size-Specific Dose Estimation; Paediatric CT Imaging Official Journal Faculty of Medicine, Universiti Sultan Zainal Abidin, Malaysia.